CALIBRATION CHAPTER 5
5-43
Annual calibration is recommended. Calibration by the user is
recorded in memory and voids factory certification. Contact your
authorized CDI sales representative for calibration and repair
services.
If extreme calibration precision is required, refer to Correction
Factors on Test Weights in this chapter.
Mounting Details
Single Transducer Mounting Detail for Calibration
Single transducer calibration requires the use of a reaction stand and
transducer calibration block.
As shown in Figure 5.1, the transducer is secured to the calibration
block with two quick release pins.
• Both pins must be installed for safety and stability of the mount.
• The white mark on the transducer is aligned in the up position.
Figure 5-1: Single Low Torque Transducer Mounting Detail